POOL: Standard control panel

Indicating the operating state of fume hoods


A control panel only supplied if the TCU-LON-II system is used as a fume hood controller

The operating state of the fume hood is indicated on the control panel. The system is configured to monitor the volume flow and / or the intake speed. This functional monitoring is stipulated by DIN EN 14175 and helps to ensure the safety of the fume hood user. If the specified conditions are not met, the alarm indicator lamp lights up and an acoustic alarm sounds.

Note:

The standard control panel can only be used in combination with certain control components.
See coding of control components
TCU-LON-II Control panel
  • Indicator LED for Alarm / operation mode / Vmax

  • Acoustic alarm

  • Indication of supply voltage failure

  • Buttons Acoustic alarm off / Vmax activation

  • Integrated LON socket for service purposes

  • Integrated LON commissing button




    • Indicator LED for Alarm (red) / Normal operation (green) / Vmax operation (yellow)

    • Failure of the supply voltage is indicated by a buffered LED.

    • Buttons for switching off acoustic alarm / activation of max. volume flow

    • Integrated LON bus access socket for service purposes

    • Integrated button function for Neuron-ID transmission during commissioning

    • Connection to the TCU-LON-II system board via a prefabricated cable (4 m)

    • Extension cables with a length of 5 m are available as an option

    • Casing dimensions approx. 23 x 150 x 19 mm for control panel installation on the fume hood housing

    • The casing can be removed for installation of the control panel into the fume cupboard housing

    • For the exact installation dimensions, see product information of the TCU-LON-II



    Name Group Type Ident-No. Type/Size
             
    TCU-LON-II control panel - User manual LABCONTROL-LON Operating Instruction E016MU1 PDF / 49 KB 

    Air management systems

    Overview brochures

    The LABCONTROL Air management systems - An introduction into system characteristics, components and typical application areas

    Installation examples

    Tailor-made solutions

    the air management system LABCONTROL

    More ...

    Contact

    TROX GmbH

    Heinrich-Trox-Platz
    D-47504 Neukirchen-Vluyn
    Tel: +49 (0)2845 / 202-0
    Contact and locations