POOL: Fume hood control and monitoring

using the LABCONTROL-LON system


Fume hoods are special safety equipment used in laboratories. Experiments that may give rise to hazardous substances are performed behind protective glass to protect the laboratory personnel.
An air flow from the laboratory into the fume hood prevents the unallowable emission of toxic matter or an increased hazardous particle concentration in the room.

Abzug-3D_240
TVLK_TMA_LON_240
TCU-LON-II_Terminal_129x480
Through non-contact and wear-free measurement of the intake speed, the exhaust volume flow is automatically adjusted as necessary to the opening width of the fume hoods protective window. It makes no difference whether the window is only partially lifted or a sash window is opened. If thermal loads are high, the exhaust volume flow is increased automatically.

This variable volume flow control in relation to the window opening position combines increased safety with lower operating costs. In addition, the lower air change rates lead to greater comfort.

The microprocessor based controller is highly precise, exceptionally fast and thus offers a high degree of safety as well as very economical cupboard operation.
All requirements for the DIN EN 14175, the British Standard and the ASHRAE are fulfilled. The examinations are performed by an independent institute and are documented in the required test certificates as per DIN EN 1475 Part 6.

Monitoring of volume flow and/or air intake speed is selectable.

In addition to the LON bus interface, the TCU-LON-II system has two additional digital inputs and one output with a changeover relay. These can be used to implement switchable alarm functions and forward alarm messages to a higher level central building control system by conventional means.

LABCONTROL fume hood controller are delivered with a standard control panel for alarm signalling and activation of special types of operation. As an option, an expanded control panel with additional functions is available.

Technical information: TCU-LON-II controller and associated components
Name Group Type Ident-No. Type/Size
         
Product overview - Fume cupboard exhaust air control LABCONTROL Technical Information 11/2008 PDF / 127 KB 
TCU-LON-II Controller LABCONTROL-LON Technical Information E016BQ2 PDF / 598 KB 
TCU-LON-II control panel - User manual LABCONTROL-LON Operating Instruction E016MU1 PDF / 49 KB 

Air management systems

Overview brochures

The LABCONTROL Air management systems - An introduction into system characteristics, components and typical application areas

Installation examples

Tailor-made solutions

the air management system LABCONTROL

More ...

Contact

TROX GmbH

Heinrich-Trox-Platz
D-47504 Neukirchen-Vluyn
Tel: +49 (0)2845 / 202-0
Contact and locations